Toronto Metropolitan University
Browse
f55cf5c667d03e2a5a5d5f3a510a7926.pdf (1022.36 kB)

Carrageenan Brain Phantom for Use in MRI

Download (1022.36 kB)
thesis
posted on 2023-12-18, 16:06 authored by Noelle Rose Crasto

Optimization of MRI is important for acquiring high-quality images during clinical examination. This can be achieved rapidly and most cost-efficiently with the use of imaging phantoms. These phantoms must be able to mimic human tissues and produce realistic relaxation times and dielectric properties. The most common MRI phantoms are constructed using either aqueous solutions or gel-like material. However, here have been few carrageenan phantoms developed for use in MRI that have shown great promise as an alternative to traditional agar phantoms. This research proposes a carrageenan brain phantom that simulates MR imaging properties as well as dielectric properties. Relaxometry and dielectric tests were performed to (1) characterize how the relaxation times and conductivity of carrageenan change with GdCl3 and NaCl. And (2) mimic white and grey brain matter using carrageenan-based phantoms.

History

Language

eng

Degree

  • Master of Engineering

Program

  • Biomedical Engineering

Granting Institution

Ryerson University

LAC Thesis Type

  • MRP

Thesis Advisor

Dafna Sussman

Year

2021

Usage metrics

    Biomedical Engineering (Theses)

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC