Version 2 2022-09-26, 14:34Version 2 2022-09-26, 14:34
Version 1 2021-05-21, 15:53Version 1 2021-05-21, 15:53
presentation
posted on 2022-09-26, 14:34authored byJ. W. Hunt, A. Y. Xuan, E. Seto, A. E. Worthington, Li Chen, Michael C. KoliosMichael C. Kolios, Michael D. Sherar
<p>Minimally invasive surgery by intense focused ultrasound
beams producing defined lesions is being studied extensively by different
groups. Lesion formation from a single pulse, depending on treatment time,
tissue temperature, and pulse repetition of about 1 minute, should produce
little damage near the skin. However, this scheme results in unacceptably long
treatment times when used on larger tumors. A possible solution is to generate
more rapid treatment times, or larger lesion volumes per pulse. However, hyperthermic
temperatures in the overlying normal tissues including the skin may limit these
treatments. In a previous presentation, simulations using an "ideal"
transducer, pulses as short as 4 s and rapid stirring of the coupling bolus
would reduce the temperature rise near the skin. Thus pulse repetitions as
short of 10 s would be acceptable. However, real transducer beams show large
aberrations which can greatly increase the near-field intensities, and make
them unacceptable for hyperthermia therapy. Some artifacts are be caused by
clamping of the transducer, others are related to thickness variations of the
transducers which generate heterogenous phase shifts from different parts of
the transducer which produce unwanted spreads at beam's focus. The authors
present detailed amplitude and phase scans near different transducers
demonstrating the artifacts, and confirm them using novel
ultrasound/magnetic-resonance phantoms showing the measured temperatures at the
focus, and at 1 cm depth from the "skin" where the heating is
considerably larger than that predicted by theory. Finally, we will discuss
solutions for problems in the near field by improving the transducer mounting
and reducing the unwanted phase shifts.</p>